Respiratory-gated helical computed tomography of lung: reproducibility of small volumes in an ex vivo model

Int J Radiat Oncol Biol Phys. 2007 Dec 1;69(5):1642-9. doi: 10.1016/j.ijrobp.2007.08.031.

Abstract

Purpose: Motion-adapted radiotherapy with gated irradiation or tracking of tumor positions requires dedicated imaging techniques such as four-dimensional (4D) helical computed tomography (CT) for patient selection and treatment planning. The objective was to evaluate the reproducibility of spatial information for small objects on respiratory-gated 4D helical CT using computer-assisted volumetry of lung nodules in a ventilated ex vivo system.

Methods and materials: Five porcine lungs were inflated inside a chest phantom and prepared with 55 artificial nodules (mean diameter, 8.4 mm +/- 1.8). The lungs were respirated by a flexible diaphragm and scanned with 40-row detector CT (collimation, 24 x 1.2 mm; pitch, 0.1; rotation time, 1 s; slice thickness, 1.5 mm; increment, 0.8 mm). The 4D-CT scans acquired during respiration (eight per minute) and reconstructed at 0-100% inspiration and equivalent static scans were scored for motion-related artifacts (0 or absent to 3 or relevant). The reproducibility of nodule volumetry (three readers) was assessed using the variation coefficient (VC).

Results: The mean volumes from the static and dynamic inspiratory scans were equal (364.9 and 360.8 mm3, respectively, p = 0.24). The static and dynamic end-expiratory volumes were slightly greater (371.9 and 369.7 mm3, respectively, p = 0.019). The VC for volumetry (static) was 3.1%, with no significant difference between 20 apical and 20 caudal nodules (2.6% and 3.5%, p = 0.25). In dynamic scans, the VC was greater (3.9%, p = 0.004; apical and caudal, 2.6% and 4.9%; p = 0.004), with a significant difference between static and dynamic in the 20 caudal nodules (3.5% and 4.9%, p = 0.015). This was consistent with greater motion-related artifacts and image noise at the diaphragm (p <0.05). The VC for interobserver variability was 0.6%.

Conclusion: Residual motion-related artifacts had only minimal influence on volumetry of small solid lesions. This indicates a high reproducibility of spatial information for small objects in low pitch helical 4D-CT reconstructions.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Artifacts
  • Lung / diagnostic imaging*
  • Models, Animal
  • Phantoms, Imaging
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Reproducibility of Results
  • Respiration*
  • Solitary Pulmonary Nodule / diagnostic imaging*
  • Swine
  • Tomography, Spiral Computed / methods*