Performance of medical radiographic X-ray systems in Greece for the time period 1998-2004

Phys Med. 2007 Dec;23(3-4):107-14. doi: 10.1016/j.ejmp.2007.10.001. Epub 2007 Nov 26.

Abstract

This study presents the results of the on-site inspections performed by the Greek Atomic Energy Commission (GAEC) on conventional X-ray systems, both in public and private medical radiology departments. A part of the inspection concerns the assessment of important radiographic parameters obtained according to a specified quality control protocol and the comparison of the measured parameter values with the corresponding acceptance limits. A total number of 1011 radiographic systems were inspected by the GAEC during the period 1998-2004, with 63.4% of them being privately owned. Analysis of 8 different operational parameters is carried out providing information on the overall performance, as well as on each parameter of the inspected X-ray systems. Tube voltage reproducibility values show the highest percentage of acceptability (98.9%, 99.5% for private and public owned radiographic systems respectively), while linearity of radiation output for private systems (72.5%) and time accuracy for public ones (72.7%) show the worst results. The comparison of the results for the private sector to those of a similar study carried out during the period 1995-1997 indicates a substantial improvement in X-ray systems performance. Higher level of improvement shows exposure time accuracy (12.2% percentile increase) and linearity of radiation output (12.5% percentile increase). Nevertheless, the situation can be further optimized if maintenance and quality control of the radiographic systems are carried out on a more regular basis.

Publication types

  • Evaluation Study

MeSH terms

  • Advisory Committees / legislation & jurisprudence
  • Greece
  • Humans
  • Private Sector
  • Public Sector
  • Radiography / methods*
  • Radiography / standards*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Time Factors