LC-MS/MS determination of 2-(4-((2-(2S,5R)-2-Cyano-5-ethynyl-1-pyrrolidinyl)-2-oxoethylamino)-4-methyl-1-piperidinyl)-4-pyridinecarboxylic acid (ABT-279) in dog plasma with high-throughput protein precipitation sample preparation

Biomed Chromatogr. 2007 Nov;21(11):1118-26. doi: 10.1002/bmc.855.

Abstract

As an effective DPP-IV inhibitor, 2-(4-((2-(2S,5R)-2-Cyano-5-ethynyl-1-pyrrolidinyl)-2-oxoethylamino)-4-methyl-1-piperidinyl)-4-pyridinecarboxylic acid (ABT-279), is an investigational drug candidate under development at Abbott Laboratories for potential treatment of type 2 diabetes. In order to support the development of ABT-279, multiple analytical methods for an accurate, precise and selective concentration determination of ABT-279 in different matrices were developed and validated in accordance with the US Food and Drug Administration Guidance on Bioanalytical Method Validation. The analytical method for ABT-279 in dog plasma was validated in parallel to other validations for ABT-279 determination in different matrices. In order to shorten the sample preparation time and increase method precision, an automated multi-channel liquid handler was used to perform high-throughput protein precipitation and all other liquid transfers. The separation was performed through a Waters YMC ODS-AQ column (2.0 x 150 mm, 5 microm, 120 A) with a mobile phase of 20 mm ammonium acetate in 20% acetonitrile at a flow rate of 0.3 mL/min. Data collection started at 2.2 min and continued for 2.0 min. The validated linear dynamic range in dog plasma was between 3.05 and 2033.64 ng/mL using a 50 microL sample volume. The achieved r(2) coefficient of determination from three consecutive runs was between 0.998625 and 0.999085. The mean bias was between -4.1 and 4.3% for all calibration standards including lower limit of quantitation. The mean bias was between -8.0 and 0.4% for the quality control samples. The precision, expressed as a coefficient of variation (CV), was < or =4.1% for all levels of quality control samples. The validation results demonstrated that the high-throughput method was accurate, precise and selective for the determination of ABT-279 in dog plasma. The validated method was also employed to support two toxicology studies. The passing rate was 100% for all 49 runs from one validation study and two toxicology studies.

Publication types

  • Evaluation Study
  • Validation Study

MeSH terms

  • Administration, Oral
  • Animals
  • Calibration / standards
  • Chemical Precipitation
  • Chromatography, High Pressure Liquid / instrumentation
  • Chromatography, High Pressure Liquid / methods*
  • Dogs
  • Drug Stability
  • Glucagon-Like Peptide 1 / blood
  • Glucagon-Like Peptide 1 / drug effects
  • Hydrophobic and Hydrophilic Interactions
  • Molecular Structure
  • Pyridines / administration & dosage
  • Pyridines / blood*
  • Pyridines / chemistry
  • Pyrrolidines / administration & dosage
  • Pyrrolidines / blood*
  • Pyrrolidines / chemistry
  • Quality Control
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Specimen Handling / methods*
  • Tandem Mass Spectrometry / methods

Substances

  • 2-(4-((2-(2-cyano-5-ethynyl-1-pyrrolidinyl)-2-oxoethyl)amino)-4-methyl-1-piperidinyl)-4-pyridinecarboxylic acid
  • Pyridines
  • Pyrrolidines
  • Glucagon-Like Peptide 1