Randomized comparative study of diamond-like carbon coated stainless steel stent versus uncoated stent implantation in patients with coronary artery disease

Arq Bras Cardiol. 2007 Apr;88(4):390-5. doi: 10.1590/s0066-782x2007000400004.
[Article in English, Portuguese]

Abstract

Objective: To compare restenosis and major cardiac event rates at one and six months after DLC-coated stent implantation with those of uncoated stents.

Methods: Randomized, prospective, comparative study of 180 patients with coronary insufficiency undergoing DLC coated stent (Phytis) or uncoated stent (Penta) implantation, from January, 2003 to July, 2004. Inclusion criteria were: de novo lesion with >50% diameter stenosis in a coronary artery with reference diameter > or = 2.5 mm and < or = 4 mm, and length < 20 mm. Exclusion criteria were: left main coronary artery and bifurcation lesions, chronic total occlusion, and in-stent restenosis.

Results: Clinical and angiographic baseline characteristics of the groups were similar. Procedural success was achieved in 98.9% of the patients in both groups. One cardiac death occurred in each group during hospitalization. Reference diameter and acute gain were greater in the Penta group (3.21 +/- 0.37 mm vs. 3.34 +/- 0.8 mm, p=0.02 and 2.3 +/- 0.5 vs. 2.49 +/- 0.5, p=0.009, respectively). Angiographic follow-up at six months showed similar rates of restenosis (24.3% vs. 21.8%, p=0.84) and of major cardiac events (16.8% vs. 17.5%, p=1).

Conclusion: DLC coated stents did not provide better outcomes in relation to uncoated stents.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Analysis of Variance
  • Angioplasty, Balloon, Coronary
  • Carbon*
  • Coated Materials, Biocompatible
  • Coronary Artery Disease / therapy*
  • Coronary Restenosis
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Stainless Steel*
  • Stents*
  • Treatment Outcome

Substances

  • Coated Materials, Biocompatible
  • Stainless Steel
  • Carbon