Determination of retention indices in constant inlet pressure mode and conversion among different column temperature conditions in comprehensive two-dimensional gas chromatography

J Chromatogr A. 2007 May 25;1150(1-2):28-36. doi: 10.1016/j.chroma.2006.09.026. Epub 2006 Sep 28.

Abstract

A method to determine the second dimensional real retention time, dead times on both dimensions and retention indices in constant inlet pressure mode was developed in comprehensive two-dimensional gas chromatography. At the same time, the conversion of GC x GC retention indices among different column temperature conditions were also conducted based on some thermodynamics parameters. The calculation accuracies are better than 1.0 retention index unit. Furthermore, a retention index database was developed and used to identify the compounds in a cigarette essential oil sample. It showed that identification by the database was of close agreement with by time-of-flight mass spectrometry, and some isomers could also be distinguished based on the retention index database.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Allergens / analysis
  • Allergens / chemistry
  • Chromatography, Gas / instrumentation*
  • Chromatography, Gas / methods*
  • Flame Ionization
  • Gas Chromatography-Mass Spectrometry
  • Perfume / analysis
  • Perfume / chemistry
  • Pressure
  • Reproducibility of Results
  • Temperature