High-performance liquid chromatographic method for the determination of mangiferin in rat plasma and urine

Biomed Chromatogr. 2006 Dec;20(12):1304-8. doi: 10.1002/bmc.695.

Abstract

A reversed-phase high-performance liquid chromatography assay for mangiferin in rat plasma and urine was developed. Rutin was employed as an internal standard. The mobile phase consisted of acetonitrile-water (16:84, v/v) containing 3% acetic acid at a flow rate of 1 mL/min. Detection was at 257 and 365 nm for mangiferin in plasma and urine, respectively. The limit of quantitation (LOQ) of mangiferin was 0.6 microg/mL in plasma, and 0.48 microg/mL in urine. The standard curve was linear from 0.6 to 24 microg/mL in plasma, and 0.48 to 24 microg/mL in urine, both intra- and inter-day precision of the mangiferin were determined and their RSD did not exceed 10%. The method provides a technique for rapid analysis of mangiferin in rat plasma and urine, which can be used in pharmacokinetic studies.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Administration, Oral
  • Animals
  • Chromatography, High Pressure Liquid / methods*
  • Male
  • Rats
  • Rats, Sprague-Dawley
  • Reproducibility of Results
  • Xanthones / administration & dosage
  • Xanthones / blood*
  • Xanthones / pharmacokinetics
  • Xanthones / urine*