Dynamics of ocular surface topography in healthy subjects

Ophthalmic Physiol Opt. 2006 Jul;26(4):419-25. doi: 10.1111/j.1475-1313.2006.00389.x.

Abstract

Our topography system is an enhancement of a standard TMS-1 corneal topograph instrument (Computed Anatomy Inc., New York, NY, USA). Topographic images are captured at a rate of 4 s(-1), allowing the recording of a series of 120 images in 30 s after a complete blink. In this prospective preliminary study 15 healthy volunteers were examined. The main outcome measures were the time profile of changes in surface regularity index (SRI), surface asymmetry index (SAI) and simulated keratometry values (K1, K2). After a blink there was a tendency for improvement in ocular surface regularity. Later trends were less clear. Our topography system makes possible the detailed evaluation of tear-film dynamics in the post-blink period. The new technique may play an important role in the diagnosis of various tear-film abnormalities; the results may also have significant implications in the planning of refractive surgeries.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Biomechanical Phenomena
  • Blinking*
  • Cornea / physiology
  • Corneal Topography / methods*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Refraction, Ocular
  • Surface Properties
  • Tears*
  • Time Factors