Accelerating MR elastography: a multiecho phase-contrast gradient-echo sequence

J Magn Reson Imaging. 2006 May;23(5):774-80. doi: 10.1002/jmri.20570.

Abstract

Purpose: To demonstrate the feasibility of using a multiecho phase-contrast (PC) gradient-echo sequence with motion-sensitizing gradient (MSG) to accelerate MR elastography (MRE) acquisitions in comparison to single-echo PC sequences.

Materials and methods: The sequence was implemented and compared with a conventional single-echo sequence as the standard of reference in both agarose phantoms and in vivo in the biceps of three healthy volunteers. For reconstruction of the elasticity modulus, a local frequency estimation (LFE) algorithm was used. ETL factors of 1-16 were evaluated.

Results: Phantom experiments demonstrated excellent consistency between single-echo and multiecho measurements in terms of wave equivalency, SNR, and reconstructed shear modulus. Additionally, the in vivo MRE examinations showed an excellent correspondence to the single-echo results. Minor loss of wave amplitude was observed at higher ETL factors.

Conclusion: The results demonstrate that a multiecho sequence is suitable for accelerating MRE in nearly homogeneous tissue, such as muscle. It provides equivalent elasticity values in a significantly reduced scan time compared to a single-echo sequence. The maximum achievable ETL factor must be individually determined for the target tissue.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acceleration
  • Adult
  • Agar
  • Algorithms
  • Elasticity
  • Feasibility Studies
  • Humans
  • Magnetic Resonance Imaging / methods*
  • Male
  • Muscle, Skeletal / anatomy & histology*
  • Muscle, Skeletal / physiology
  • Phantoms, Imaging
  • Shear Strength
  • Stress, Mechanical
  • Vibration