Processing gas chromatographic data and confidence interval calculation for partition coefficients determined by the phase ratio variation method

J Chromatogr A. 2006 Mar 31;1110(1-2):146-55. doi: 10.1016/j.chroma.2006.01.055. Epub 2006 Jan 31.

Abstract

The phase ratio variation (PRV) method is widely used for the determination of partition coefficient values (dimensionless Henry's law constants) by headspace gas chromatography. Traditional data processing by linear regression has several drawbacks: potential bias introduced by linearization, absence of quality indicator of the resulting value and, in case of replicate determinations, poor utilisation of the existing measurements leading to unnecessarily large confidence intervals. The paper compares existing PRV data processing methods (linear and nonlinear regression, parametric) and derives confidence intervals for the resulting partition coefficient values. The possibility of using several series of measurements to derive a single partition coefficient value with tighter and more reliable confidence intervals is presented for all three processing methods. The methods are tested on published literature data and new experimental data for 12 volatile organic compounds in water at 25 degrees C. The nonlinear regression based on several series of measurements appears to be the method of choice.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, Gas / methods*
  • Confidence Intervals*
  • Mathematics
  • Organic Chemicals / chemistry*
  • Pharmaceutical Preparations / analysis*
  • Regression Analysis
  • Solubility
  • Temperature
  • Volatilization

Substances

  • Organic Chemicals
  • Pharmaceutical Preparations