Implementation, accuracy evaluation, and preliminary clinical trial of a CT-free navigation system for high tibial opening wedge osteotomy

Comput Aided Surg. 2005 Mar;10(2):73-85. doi: 10.3109/10929080500228837.

Abstract

Objective: The objectives of this study are to design and evaluate a CT-free intra-operative planning and navigation system for high tibial opening wedge osteotomy. This is a widely accepted treatment for medial compartment osteoarthritis and other lower extremity deformities, particularly in young and active patients for whom total knee replacement is not advised. However, it is a technically demanding procedure. Conventional preoperative planning and surgical techniques have so far been inaccurate, and often resulting in postoperative malalignment representing either under- or over-correction, which is the main reason for poor long-term results. In addition, conventional techniques have the potential to damage the lateral hinge cortex and tibial neurovascular structures, which may cause fixation failure, loss of correction, or peroneal nerve paralysis. All these common problems can be addressed by the use of a surgical navigation system.

Materials and methods: Surgical instruments are tracked optically with the SurgiGATE((R)) navigation system (PRAXIM MediVision, La Tronche, France). Following exposure, dynamical reference bases are attached to the femur, tibia, and proximal fragment of the tibia. A patient-specific coordinate system is then established, on the basis of registered anatomical landmarks. After intra-operative deformity measurement and correction planning, the osteotomy is performed under navigational guidance. The deformities are corrected by realigning the mechanical axis of the affected limb from the diseased medial compartment to the healthy lateral side. The wedge size, joint line orientation, and tibial plateau slope are monitored during correction. Besides correcting uni-planar varus deformities, the system provides the functionality to correct complex multi-planar deformities with a single cut. Furthermore, with on-the-fly visualization of surgical instruments on multiple fluoroscopic images, penetration of the hinge cortex and damage to the neurovascular structures due to an inappropriate osteotomy can be avoided.

Results: The laboratory evaluation with a plastic bone model (Synbone AG, Davos, Switzerland) shows that the error of deformity correction is <1.7 degrees (95% confidence interval) in the frontal plane and <2.3 degrees (95% confidence interval) in the sagittal plane. The preliminary clinical trial confirms these results.

Conclusion: A novel CT-free navigation system for high tibial osteotomy has been developed and evaluated, which holds the promise of improved accuracy, reliability, and safety of this procedure.

Publication types

  • Clinical Trial
  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Arthroscopy / methods*
  • Fluoroscopy
  • Humans
  • Image Interpretation, Computer-Assisted
  • Knee Joint / diagnostic imaging
  • Knee Joint / surgery
  • Minimally Invasive Surgical Procedures / instrumentation
  • Minimally Invasive Surgical Procedures / methods
  • Monitoring, Intraoperative / instrumentation*
  • Observer Variation
  • Osteotomy / instrumentation*
  • Osteotomy / methods
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Surgery, Computer-Assisted / methods*
  • Technology Assessment, Biomedical
  • Tibia / surgery*
  • Tomography, X-Ray Computed