[Study on electrochemical mechanism of coronary stent used austenitic stainless steel in flowing artificial body fluid]

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2005 Aug;22(4):730-3.
[Article in Chinese]

Abstract

The electrochemical mechanism of austenitic stainless steel (SUS316L and SUS317L) coronary stents in flowing artificial body fluid has been investigated with electrochemical technologies. The results indicated that the flowing medium coursed the samples' pitting potential Eb shift negatively, increased the pitting corrosion sensitivity, accelerated its anodic dissolution, but had little effects on repassivated potential. The flowing environment had great effects on cathodic process. The oxygen reaction on the samples' surface became faster as the cathodic process was not controlled by oxygen diffusion but by mixed diffusion and electrochemical process. With the increase of velocity of solution, the pitting corrosion becomes liable to occur under this circumstance.

Publication types

  • English Abstract
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biocompatible Materials / chemistry*
  • Body Fluids / chemistry
  • Corrosion
  • Electrochemistry
  • Materials Testing
  • Stainless Steel / chemistry*
  • Stents*

Substances

  • Biocompatible Materials
  • Stainless Steel