A comparison of a similarity-based and a feature-based 2-D-3-D registration method for neurointerventional use

IEEE Trans Med Imaging. 2005 Aug;24(8):1058-66. doi: 10.1109/TMI.2005.852067.

Abstract

Two-dimensional (2-D)-to-three-dimensional (3-D) registration can improve visualization which may aid minimally invasive neurointerventions. Using clinical and phantom studies, two state-of-the-art approaches to rigid registration are compared quantitatively: an intensity-based algorithm using the gradient difference similarity measure; and an iterative closest point (ICP)-based algorithm. The gradient difference approach was found to be more accurate, with an average registration accuracy of 1.7 mm for clinical data, compared to the ICP-based algorithm with an average accuracy of 2.8 mm. In phantom studies, the ICP-based algorithm proved more reliable, but with more complicated clinical data, the gradient difference algorithm was more robust. Average computation time for the ICP-based algorithm was 20 s per registration, compared with 14 min and 50 s for the gradient difference algorithm.

Publication types

  • Clinical Trial
  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Arteriovenous Malformations / diagnosis
  • Artificial Intelligence
  • Cluster Analysis
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / methods*
  • Intracranial Aneurysm / diagnosis
  • Neuronavigation / methods*
  • Pattern Recognition, Automated / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Subtraction Technique*