A semiautomated ultrasound border detection program that facilitates clinical measurement of ultrasound carotid intima-media thickness

J Am Soc Echocardiogr. 2005 Mar;18(3):244-51. doi: 10.1016/j.echo.2004.12.002.

Abstract

We have developed a novel, semiautomated carotid intima-media thickness (CIMT) border detection program (AUTO) and evaluated its measurement reproducibility and accuracy. Images from 6 carotid segments were acquired in 50 subjects, for a total of 300 segments. Mean and maximum CIMT values were measured blindly at a reference (REF) lab and in duplicate by experienced (EXP) and novice (NOV) readers using manual (MAN) and AUTO methods. Coefficients of variation for AUTO measurements of mean (3.2%) and maximum (4.1%) CIMT were low, and the AUTO method improved the NOV reader's reproducibility. Compared with the REF lab, mean (0.012 +/- 0.006 mm) and maximum (0.144 +/- 0.006 mm) CIMT biases were small and equivalent to those of the REF lab ( P < .001). The AUTO method shortened reading times by 35% to 46% ( P < .001). We conclude that our novel AUTO CIMT measurement program improved reproducibility and was accurate. Compared with MAN tracing, the AUTO method agreed better with the REF lab and decreased reading time.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adult
  • Automation
  • Carotid Artery Diseases / diagnostic imaging*
  • Carotid Artery, Common / diagnostic imaging*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Reproducibility of Results
  • Tunica Intima / diagnostic imaging*
  • Tunica Media / diagnostic imaging*
  • Ultrasonography / instrumentation
  • Ultrasonography / methods*