Extremely low frequency magnetic fields originating from equipment used for assisted reproduction, umbilical cord and peripheral blood stem cell transplantation, transfusion, and hemodialysis

Bioelectromagnetics. 2005 Jan;26(1):69-73. doi: 10.1002/bem.20071.

Abstract

The extremely low-frequency (ELF) magnetic fields (MFs) originating from equipment used for assisted reproduction, umbilical cord-blood and peripheral-blood stem cell transplantation, transfusion, and hemodialysis were measured. The ELF-MF values were 0.1-1.2 microT on clean benches, <0.1-8.0 microT on inverted microscopes, <0.1-13.6 mmicroT in CO2 incubators, 4.3-11.5 microT in centrifuges, 0.4-18.8 microT in programmed freezers, <0.1-0.3 microT in deep freezers, 0.3-3.1 microT on cell separators, and 0.2-0.9 microT in hemodialysers. Frequencies of MFs were nominally 60 Hz, but some devices showed non-sinusoidal 120 Hz. Such MFs can be reduced by shielding the sources or altering the protocols employed.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Transfusion / instrumentation
  • Cord Blood Stem Cell Transplantation / instrumentation
  • Equipment Failure Analysis / methods*
  • Equipment Safety / methods*
  • Equipment and Supplies*
  • Japan
  • Peripheral Blood Stem Cell Transplantation / instrumentation
  • Radiation Dosage
  • Radiation Protection / methods*
  • Radiometry / methods*
  • Renal Dialysis / methods
  • Reproductive Techniques, Assisted / instrumentation
  • Risk Assessment / methods*
  • Risk Factors