In vitro exposure apparatus for ELF magnetic fields

Bioelectromagnetics. 2004 Dec;25(8):582-91. doi: 10.1002/bem.20037.

Abstract

For in vitro studies on the effect of extremely low frequency (ELF) magnetic field exposures in different laboratories, a programmable, high precision exposure system enabling blinded exposures has been developed and fully characterized. It is based on two shielded 4 coil systems that fit inside a commercial incubator. The volume of uniform B field exposure with 1% field tolerance is 50% larger compared to a Merrit 4 coil system with the same coil volume. The uncertainties for the applied magnetic fields have been specified to be less than 4%. The computer controlled apparatus allows signal waveforms that are composed of several harmonics, blind protocols, monitoring of exposure and environmental conditions and the application of B fields up to 3.6 mT root-mean-square amplitude. Sources of artifacts have been characterized: sham isolation >43 dB, parasitic incident E fields <1 V/m, no recognizable temperature differences in the media for exposure or sham state, and vibrations of the mechanically decoupled dish holder <0.1 m/s(2) (= 0.01 g), which is only twice the sham acceleration background level produced by the incubator and fan vibrations.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cell Culture Techniques / instrumentation*
  • Cell Culture Techniques / methods
  • Electromagnetic Fields
  • Electromagnetic Phenomena / instrumentation*
  • Electromagnetic Phenomena / methods
  • Environment, Controlled
  • Equipment Design
  • Equipment Failure Analysis
  • Radiation Dosage
  • Radiometry / instrumentation*
  • Transducers*