[Comparitive study of the threshold variability between blue-on-yellow perimetry and white-on-white perimetry]

Yan Ke Xue Bao. 2002 Jun;18(2):87-91.
[Article in Chinese]

Abstract

Purpose: To compare the threshold variability of blue-on-yellow (B/Y) perimetry with white-white(W/W) perimetry, and evaluate the reproducibility of B/Y perimetry.

Methods: The B/Y perimetry and W/W perimetry in the Octopus 101 perimetry were used to examine the visual fields of 12 normal subjects (24 eyes), 16 cases (32 eyes) of primary open angle glaucoma (POAG), and 7 cases (14 eyes) of suspected POAG respectively. The B/Y perimetry and W/W perimetry were repeated to examine within 2 weeks. The point by point threshold variability of the two perimetries were compared and analysed.

Results: The total mean threshold variability for B/Y perimetry (2.61 +/- 0.94) dB was greater than that for W/W perimetry (2.11 +/- 0.90) dB in all subjects, but it had no significance (P = 0.6244). The total mean threshold variability for B/Y perimetry (2.07 +/- 0.54) dB was significantly greater than that for W/W perimetry (1.50 +/- 0.34) dB in normal subjects (P = 0.0006), while no significance was found in suspected POAG and POAG groups between the two perimetries (P = 0.0523 and 0.9371). The threshold variability in the areas of some eccentricities for B/Y perimetry were significantly greater than that for W/W perimetry in normal subjects and suspected POAG group, but no significance was found in POAG group for all eccentricities between the two perimetries.

Conclusion: The threshold variability for B/Y perimetry was greater than that for W/W perimetry in normal subjects, but no significance was found in suspected POAG and POAG between the two perimetries.

Publication types

  • Comparative Study
  • English Abstract

MeSH terms

  • Adult
  • Female
  • Glaucoma, Open-Angle / diagnosis*
  • Humans
  • Male
  • Middle Aged
  • Sensitivity and Specificity
  • Sensory Thresholds
  • Visual Field Tests / methods*
  • Visual Perception*