Uniform illumination of optically dense NMR samples

J Magn Reson. 2004 Nov;171(1):171-5. doi: 10.1016/j.jmr.2004.08.017.

Abstract

We demonstrate a simple, inexpensive method for in situ laser illumination of NMR samples using a stepwise tapered optical fibre to deliver light uniformly along the axis of a 5 mm NMR tube. The optical path length of the incident light inside the sample is about 3 mm, allowing efficient illumination of optically dense samples. The degradation in spectral resolution and the reduction in filling factor are both minimal. Probe modifications are not required.