In vitro host response assessment of biomaterials for cardiovascular stent manufacture

J Mater Sci Mater Med. 2004 Apr;15(4):473-7. doi: 10.1023/b:jmsm.0000021123.51752.11.

Abstract

The deployment of a vascular stent during angioplasty has greatly reduced the risks of restenosis. However, the presence of the device still induces a host response as well as a mechanical action on the blood vessel wall and an alteration of the haemodynamics. Platelet and inflammatory cells can adhere on the stent surface and be activated to produce biochemical signals able to stimulate an excessive proliferation of the smooth muscle cells with the consequent obstruction of the vessel lumen. For these reasons, the host response to two of the materials used in stent manufacture, stainless steel and diamond-like carbon, was investigated in vitro. The data showed that stainless steel induced a higher level of host response both in terms of platelet aggregation and macrophage activation. However, the spreading of inflammatory cells was more accentuated on diamond-like carbon. The inflammatory cells produced levels of platelet-derived growth factor, a key signal in smooth muscle cell proliferation, similar to stainless steel thus suggesting that carbon coatings may not be able to prevent restenosis.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adsorption
  • Biocompatible Materials / chemistry
  • Blood Proteins / chemistry
  • Blood Proteins / metabolism
  • Blood Vessel Prosthesis / adverse effects*
  • Cells, Cultured
  • Diamond / chemistry*
  • Foreign-Body Reaction / etiology*
  • Foreign-Body Reaction / immunology*
  • Foreign-Body Reaction / metabolism
  • Foreign-Body Reaction / pathology
  • Humans
  • Leukocytes, Mononuclear / immunology
  • Leukocytes, Mononuclear / pathology
  • Macrophage Activation / immunology
  • Materials Testing / methods*
  • Platelet Adhesiveness / immunology
  • Platelet Aggregation / immunology
  • Stainless Steel / chemistry*
  • Stents / adverse effects*
  • Surface Properties

Substances

  • Biocompatible Materials
  • Blood Proteins
  • Stainless Steel