Abstract

A high-performance liquid chromatographic (HPLC) method with fluorescence detection has been developed for the determination of rizatriptan in human plasma. Following a single-step liquid-liquid extraction with methyl tertiarybutyl ether, the analytes were separated using a mobile phase consisting of 0.05% (v/v) triethylamine in water (adjusting to pH 2.75 with 85% phosphoric acid) and acetonitrile (92:8, v/v). Fluorescence detection was performed at an excitation wavelength of 225nm and an emission wavelength of 360nm. The linearity for rizatriptan was within the concentration range of 0.5-50ng/ml. The intra- and inter-day precisions of the method were not more than 8.0%. The lower limit of quantification (LLOQ) was 0.5ng/ml for rizatriptan. The method was sensitive, simple and repeatable enough to be used in pharmacokinetic studies.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Serotonin Receptor Agonists / blood*
  • Serotonin Receptor Agonists / pharmacokinetics
  • Spectrometry, Fluorescence
  • Triazoles / blood*
  • Triazoles / pharmacokinetics
  • Tryptamines

Substances

  • Serotonin Receptor Agonists
  • Triazoles
  • Tryptamines
  • rizatriptan