[Experimental study of diode-laser induced thermocoagulation on hepatic tissue with scanner fiber tip]

Zhonghua Wai Ke Za Zhi. 2003 Nov;41(11):849-51.
[Article in Chinese]

Abstract

Objective: To seek a safe, efficient, and cost-effective technique for local thermo-ablation of hepatic cancer.

Methods: The livers from 16 healthy rabbits were thermocoagulated by diode-laser with scanner fiber tip, 6 w for 10 mins. At the same time, the temperatures were measured at 0, 5 and 10 mm from laser tip. The pre-thermocoagulative liver function was compared with that of 7 days post-thermocoagulation. The pathologic changes were also observed 1 month after laser thermocoagulation.

Results: All the rabbits survived and hepatic tissue temperatures at 0, 5, 10 mm from laser tip reached 96.39 degrees C +/- 3.97 degrees C, 60.79 degrees C +/- 6.21 degrees C, 46.10 degrees C +/- 4.58 degrees C respectively after 10 minutes of thermocoagulation. There were no significant differences in liver function parameters between rabbits of pre-laser thermocoagulation and of post-laser thermocoagulation. Thermocoagulated necrosis of liver tissue with surrounding fibrosis in a diameter of 26.0 mm was formed. Light microscopy revealed coagulative necrosis in the center of the coagulated area without surviving hepatic cells.

Conclusion: The hepatic tissue can be coagulated safely and effectively by diode-laser with scanner fibertip, and such a technique may provide a new method for the treatment of hepatic carcinoma.