Determination of bopindolol using the flow injection technique coupled with solid phase extraction

J Pharm Biomed Anal. 2003 Dec 4;33(5):1149-53. doi: 10.1016/s0731-7085(03)00430-8.

Abstract

In the proposed procedure, the determination of bopindolol using a flow injection analysis (FIA) technique, with spectrophotometric detection at 635 nm, is described. The method is based on the production of a green, water-soluble complex with ferric ions in acid medium. The automated lab-made FIA system was used for the direct determination of bopindolol in tablets. Bopindolol was adsorbed onto the solid phase in a mini-column, which was integrated directly into the flow system. The positive feature of the use of solid phase extraction (SPE) was the pre-concentration of bopindolol (seven times). The sample throughput was 50 samples per hour. Using the SPE method, bopindolol was determined with a linear range from 125 to 1000 microg ml(-1) (Relative standard deviation (R.S.D.)=1.87%), with a detection limit (3sigma) of 70 microg ml(-1). The method was applied to the determination of bopindolol in Sandonorm tablets. The results obtained were compared with a conventional HPLC method, both analytical techniques were in good agreement.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Flow Injection Analysis / instrumentation
  • Flow Injection Analysis / methods
  • Pindolol / analogs & derivatives*
  • Pindolol / analysis*
  • Pindolol / chemistry
  • Tablets

Substances

  • Tablets
  • Pindolol
  • bopindolol