Reliable method for the determination of ranitidine by liquid chromatography using a microvolume of plasma

J Chromatogr B Analyt Technol Biomed Life Sci. 2003 Sep 25;795(1):141-4. doi: 10.1016/s1570-0232(03)00474-4.

Abstract

The aim of the present study was to develop a simple method to measure ranitidine, using 100 microL of plasma, by high-performance liquid chromatography with a Symmetry C(18) column and UV detection at 313 nm. Linearity was assessed in the range from 50 to 1500 ng ml(-1) and had a correlation coefficient of 0.999. The inter- and intra-day coefficients of variation were less than 7%. The limits of detection and quantitation were 5 and 15 ng ml(-1), respectively. Drug levels were determined satisfactorily in three patients. A simple and reliable method was developed which uses a microvolume of plasma, particularly useful in low-weight children.

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Histamine H2 Antagonists / blood*
  • Humans
  • Ranitidine / blood*
  • Reproducibility of Results
  • Spectrophotometry, Ultraviolet / methods

Substances

  • Histamine H2 Antagonists
  • Ranitidine