[Validity of the proximal isovelocity surface area color Doppler method for calculating the valve area in patients with mitral stenosis; comparison with the two-dimensional echocardiographic method]

G Ital Cardiol. 1992 Oct;22(10):1201-10.
[Article in Italian]

Abstract

Background: The proximal isovelocity surface area (PISA) method, assessed by color Doppler echocardiography, has gained acceptance as a means of calculating flow rate through regurgitant orifice. The method can also be used to derive mitral valve area (MVA), by continuity equation, in patients with mitral stenosis (MS). The aim of this study was to compare the PISA method with the two-dimensional echocardiographic planimetry (2D) method and pressure half-time method (PHT) in MVA calculations in a group of 37 patients with MS.

Methods and results: All of these patients had satisfactory MVA by 2D method. There were 22 female and 15 male; age 56 +/- 11 years (range 32-71); 19 were in sinus rhythm (SR) and 18 in atrial fibrillation (AF); 17 patients had pure MS, while the remaining 20 had associated mitral regurgitation (MR); in 23 patients the orifice morphology was circular or elliptic, and was defined as regular; while in 14 patients the morphology was irregular for the presence of two or more nodular calcifications on the commissures or leaflet's edges. MVA by PISA method was calculated assuming a uniform radial flow convergence region along a hemispherical surface, according to the formula: MVA = 2 pi r2 Vn(1-cos theta)/Vmax; where r was the PISA radius measured in 2D from the first alias to the mitral leaflet's edge; Vn was the flow velocity at radial distance from the mitral orifice; Vmax was the peak transmitral velocity by CW Doppler; 1-cos theta was a factor that accounted for the inflow angle formed by the mitral leaflets. The Nyquist limit was lowered to 29 cm/sec. Alpha angle formed by the mitral leaflets ranged between 86 degrees and 134 degrees; average 110 degrees +/- 10 degrees. 2D MVA was 1.33 +/- 0.37 cm2; range 0.69-2.2 cm2; PHT MVA was 1.29 +/- 0.34 cm2; range 0.70-2.1 cm2; PISA MVA was 1.18 +/- 0.36 cm2; range 0.47-1.95 cm2. The PISA method underestimates MVA by 0.15 +/- 0.21 cm2, in comparison with the 2D method; and by 0.11 +/- 0.18 cm2 in comparison with PHT method (p ns). The correlation between 2D and PISA MVA was: r = 0.84; p < 0.001; y = 0.83x + 0.06; 95% confidence intervals +/- 0.40 cm2; and between PHT and PISA MVA was: r = 0.79; y = 0.84x + 0.09; p < 0.001; 95% confidence intervals +/- 0.46 cm2. The correlation coefficient was similarly good in patients with SR or AF, and did not significantly change in patients with pure MS or MS+MR; neither did it vary with respect to the orifice morphology (p < 0.001 for all the variables considered), except for the correlation PHT-PISA in the group of patients with irregular orifice morphology (r = 0.70; p = 0.005). The interobserver and intraobserver variability were, respectively: 2.2% and 4.4% for 2D MVA; 3.4% and 3.8% for PHT MVA; 5.2% and 3.5% for the PISA radius; 6.1% and 4.4% for the alpha angle; 10.2% and 7.2% for PISA MVA (F ratio of variances ns).

Conclusions: In conclusion, the PISA method allows accurate assessment of MVA in patients with MS, regardless of cardiac rhythm or additional MR. Moreover, our study suggests that orifice morphology does not affect the accuracy of this method.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Echocardiography, Doppler / methods*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Mitral Valve / diagnostic imaging*
  • Mitral Valve / pathology
  • Mitral Valve Stenosis / diagnostic imaging*
  • Mitral Valve Stenosis / pathology
  • Mitral Valve Stenosis / physiopathology