An automatic flow procedure for the determination of 3-hydroxybutyrate in animal serum and plasma

J Agric Food Chem. 2003 Apr 23;51(9):2457-60. doi: 10.1021/jf021024n.

Abstract

An automatic flow procedure based on the multicommutation concept, comprising three-way solenoid valves, for the spectrophotometric determination of 3-hydroxybutyrate in animal serum and plasma is proposed. The 3-hydroxybutyrate was enzymatically converted to acetoacetate with the reduction of NAD+ to NADH monitored at 340 nm. It was possible to carry out up to 600 determinations without a significant decrease in the analytical signal, with 5 mg of 3-hydroxybutyrate dehydrogenase immobilized on porous silica beads and packed in a column. The system enabled 60 determinations/h of 3-hydroxybutyrate in the range of 10-150 mg L(-1), with a consumption of 0.9 mg of NAD+ and 200 microL of sample per determination. A detection limit of 2 mg L(-1) for both animal serum and plasma and coefficients of variation of 1.4% and 1.2% (n = 17), respectively, were determined. Animal serum and plasma samples were analyzed without previous treatment, the results of which agreed with those obtained using the conventional method (UV kit, Sigma).

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • 3-Hydroxybutyric Acid / blood*
  • 3-Hydroxybutyric Acid / metabolism
  • Animals
  • Cattle / blood*
  • Hydrogen-Ion Concentration
  • Hydroxybutyrate Dehydrogenase / metabolism
  • Male
  • Oxidation-Reduction
  • Sensitivity and Specificity
  • Sheep / blood*
  • Spectrophotometry / methods*
  • Temperature

Substances

  • Hydroxybutyrate Dehydrogenase
  • 3-Hydroxybutyric Acid