Simultaneous micellar LC determination of lidocaine and tolperisone

J Pharm Biomed Anal. 2003 Mar 26;31(4):629-38. doi: 10.1016/s0731-7085(02)00693-3.

Abstract

A micellar liquid chromatography (MLC) procedure was developed for the simultaneous separation and determination of lidocaine hydrochloride (LD HCl) and tolperisone hydrochloride (TP HCl) using a short-column C18 (12.5 mm x 4.6 mm, 5 microm), sodium dodecyl sulfate (SDS) with a small amount of isopropanol, and diode array detector. The optimum conditions for the simultaneous determination of both drugs were 0.075 mol l(-1) SDS-7.5% (v/v) isopropanol with a flow rate of 0.7 ml min(-1) and detection at 210 nm. The LOD (2S/N) of LD HCl was 0.73 ng 20 microl(-1), whereas that of TP HCl was 1.43 ng 20 microl(-1). The calibration curves for LD HCl and TP HCl were linear over the ranges 0.125-500 microg ml(-1) (r(2)=0.9999) and 1.00-500 microg ml(-1) (r(2)=0.9997), respectively. The %recoveries of both drugs were in the range 98-103% and the %RSD values were less than 2. The proposed method has been successfully applied to the simultaneous determination of TP HCl and LD HCl in various pharmaceutical preparations.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Chromatography, Liquid / methods
  • Chromatography, Micellar Electrokinetic Capillary / methods*
  • Lidocaine / analysis*
  • Tolperisone / analysis*

Substances

  • Lidocaine
  • Tolperisone