Loss of left ventricular epicardial lead capture due to pneumothorax

Pacing Clin Electrophysiol. 2002 Jun;25(6):996-7. doi: 10.1046/j.1460-9592.2002.00996.x.

Abstract

Pneumothorax is a rare complication of pacemaker lead insertion by subclavian way. We report a case of temporary ineffective biventricular pacing due to pneumothorax. This complication has to be ruled out before electing to reposition or to replace the lead.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Bradycardia / therapy*
  • Equipment Failure
  • Female
  • Heart Ventricles
  • Humans
  • Pacemaker, Artificial / adverse effects*
  • Pneumothorax / etiology*
  • Pneumothorax / physiopathology