Detection of osteoporosis by dual energy X-ray absorptiometry (DXA) of the calcaneus: is the WHO criterion applicable?

Calcif Tissue Int. 2002 Jun;70(6):475-82. doi: 10.1007/s002230020030.

Abstract

The study assessed the precision, sensitivity, and specificity of a recently developed peripheral dual-energy X-ray absorptiometry (DXA) scanner, applied to the calcaneus, in the identification of individuals with osteoporosis at axial sites by DXA. Two hundred and two women, aged 55.2 +/- l3.7 years (mean +/- SD), participated in the study. The precisions (coefficient of variation) of measurements in vitro (0.48%) and in vivo (1.40%) were very good. The in vivo precision was independent of the operator, foot size, foot width, weight, height, and body mass index. Calcaneus BMD correlated moderately (r = 0.494-0.690, P <0.001) with axial BMD measurements by DXA. Using the World Health Organization (WHO) criterion for defining osteoporosis (T score ? -2.5) the specificity of the calcaneus to identify patients with osteoporosis at total hip, femoral neck, spine, or any of these axial sites was excellent (97.0%, 97.0%, 96.5%, and 97.1%, respectively); however, the sensitivity was poor (58.8%, 36.4%, 21.8%, and 20.3%, respectively). Therefore, the WHO criterion is not appropriate for DXA calcaneus. Based on femoral neck BMD for detection of osteoporosis, a more appropriate calcaneus T score threshold would be -1.4 by analyses of receiver-operator characteristic curves; this might serve to select those patients who might appropriately be referred for axial DXA.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorptiometry, Photon*
  • Adult
  • Aged
  • Aged, 80 and over
  • Bone Density*
  • Calcaneus / diagnostic imaging*
  • Female
  • Femur Neck / diagnostic imaging
  • Hip Joint / diagnostic imaging
  • Humans
  • Lumbar Vertebrae / diagnostic imaging
  • Middle Aged
  • Osteoporosis, Postmenopausal / classification
  • Osteoporosis, Postmenopausal / diagnosis*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • World Health Organization*