k-space filtering in 2D gradient-echo breath-hold hyperpolarized 3He MRI: spatial resolution and signal-to-noise ratio considerations

Magn Reson Med. 2002 Apr;47(4):687-95. doi: 10.1002/mrm.10134.

Abstract

In this work some of the factors that can influence the signal-to-noise ratio (SNR) and spatial resolution in MR images of inhaled hyperpolarized gases are systematically addressed. In particular, the effects of RF depletion of longitudinal polarization and image gradient diffusion dephasing were assessed in terms of their contribution to a k-space filter. By means of theoretical simulations and a novel method of experimental validation using a variable transverse magnetization of the 1H signal, systematic quantitative and qualitative investigations of the effects of k-space filtering intrinsic to imaging of hyperpolarized gas were made. A 2D gradient-echo image is considered for a range of flip angles with centric, sequential, and half-Fourier Cartesian phase-encoding strategies, and the results are assessed in terms of SNR and spatial resolution in the reconstructed images. Centric phase encoding was found to give the best SNR at higher flip angles, with a trade-off in spatial resolution compared to sequential phase encoding. A half-Fourier approach potentially offers increased SNR through the use of higher flip angles without compromising the spatial resolution, which is comparable to that achieved with sequential encoding.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Fourier Analysis
  • Helium
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Isotopes
  • Magnetic Resonance Imaging / methods*