Biosensor arrays for simultaneous measurement of glucose, lactate, glutamate, and glutamine

Biosens Bioelectron. 2002 Apr;17(4):297-302. doi: 10.1016/s0956-5663(01)00298-6.

Abstract

For simultaneous measurement of glucose, lactate, glutamine, and glutamate a biosensor array is implemented in a micro flow-system thus giving a microsystem. The microsystem consists of a glass chip with the integrated biosensor array and a bottom part, which comprises a gold counter electrode, a 300 microm thick seal, and electrical interconnection lines. The flow device has a total internal volume of 2.1 or 6 microl when integrated with a mixer on chip. The biosensors with no crosstalking and high long term stability were produced by modifying the electrochemical transducers and utilizing photopatternable enzyme membranes. The use of appropriate miniaturization technology leads to mass producable devices for in vivo and ex vivo applications in whole blood and fermentation broth. Due to a novel glutaminase with an activity optimum in the neutral pH range direct and simultaneous monitoring of glutamine together with glucose, lactate, and glutamate could be performed.

MeSH terms

  • Biosensing Techniques / instrumentation*
  • Biosensing Techniques / methods
  • Enzymes, Immobilized / metabolism
  • Equipment Design
  • Glucose / analysis*
  • Glucose / metabolism
  • Glutamic Acid / analysis*
  • Glutamic Acid / metabolism
  • Glutamine / analysis*
  • Glutamine / metabolism
  • Lactic Acid / analysis*
  • Lactic Acid / metabolism
  • Membranes, Artificial
  • Microelectrodes
  • Polyhydroxyethyl Methacrylate / chemistry

Substances

  • Enzymes, Immobilized
  • Membranes, Artificial
  • Glutamine
  • Polyhydroxyethyl Methacrylate
  • Lactic Acid
  • Glutamic Acid
  • Glucose