HLA-B typing by reference strand mediated conformation analysis using a capillary-based semiautomated genetic analyzer

Hum Immunol. 2001 Apr;62(4):414-8. doi: 10.1016/s0198-8859(01)00213-0.

Abstract

The application of reference strand conformation analysis (RSCA) to HLA-A typing using the ABI PRISM 310 capillary based genetic analyzer has recently been described. This study outlines the development and validation of capillary RSCA for HLA-B typing. Mobility values for 93 HLA-B alleles were defined following electrophoresis of known controls through the system. Three fluorescently labelled references, labelled with three different dyes can be electrophoresed simultaneously. The technique was validated by comparing results from 296 cord blood donors with those obtained using reverse SSO. Following capillary RSCA 14.5% of samples required confirmatory typing, compared with a repeat rate of 5.1% following reverse SSO. In samples where no other typing was necessary there was 100% correlation between the two methods. Capillary RSCA for HLA-B typing is quick, easy to implement, and with the introduction of new FLRs and gel matrices has the potential to evolve into a high resolution typing method.

MeSH terms

  • Alleles
  • Automation
  • HLA-B Antigens / classification
  • HLA-B Antigens / genetics*
  • Histocompatibility Testing / methods*
  • Humans
  • Nucleic Acid Conformation*
  • Polymerase Chain Reaction / methods*

Substances

  • HLA-B Antigens