Commissioning an image-guided localization system for radiotherapy

Int J Radiat Oncol Biol Phys. 2000 Aug 1;48(1):267-76. doi: 10.1016/s0360-3016(00)00581-2.

Abstract

Purpose: To describe the design and commissioning of a system for the treatment of classes of tumors that require highly accurate target localization during a course of fractionated external-beam therapy. This system uses image-guided localization techniques in the linac vault to position patients being treated for cranial tumors using stereotactic radiotherapy, conformal radiotherapy, and intensity-modulated radiation therapy techniques. Design constraints included flexibility in the use of treatment-planning software, accuracy and precision of repeat localization, limits on the time and human resources needed to use the system, and ease of use.

Methods and materials: A commercially marketed, stereotactic radiotherapy system, based on a system designed at the University of Florida, Gainesville, was adapted for use at the University of Washington Medical Center. A stereo pair of cameras in the linac vault were used to detect the position and orientation of an array of fiducial markers that are attached to a patient's biteblock. The system was modified to allow the use of either a treatment-planning system designed for stereotactic treatments, or a general, three-dimensional radiation therapy planning program. Measurements of the precision and accuracy of the target localization, dose delivery, and patient positioning were made using a number of different jigs and devices. Procedures were developed for the safe and accurate clinical use of the system.

Results: The accuracy of the target localization is comparable to that of other treatment-planning systems. Gantry sag, which cannot be improved, was measured to be 1.7 mm, which had the effect of broadening the dose distribution, as confirmed by a comparison of measurement and calculation. The accuracy of positioning a target point in the radiation field was 1.0 +/- 0.2 mm. The calibration procedure using the room-based lasers had an accuracy of 0.76 mm, and using a floor-based radiosurgery system it was 0.73 mm. Target localization error in a phantom was 0.64 +/- 0.77 mm. Errors in positioning due to couch rotation error were reduced using the system.

Conclusion: The system described has proven to have acceptable accuracy and precision for the clinical goals for which it was designed. It is robust in detecting errors, and it requires only a nominal increase in setup time and effort. Future work will focus on evaluating its suitability for use in the treatment of head-and-neck cancers not contained within the cranial vault.

MeSH terms

  • Brain Neoplasms / radiotherapy*
  • Calibration
  • Equipment Design
  • Evaluation Studies as Topic
  • Humans
  • Particle Accelerators
  • Phantoms, Imaging
  • Physical Phenomena
  • Physics
  • Radiosurgery / instrumentation*
  • Radiotherapy Planning, Computer-Assisted / instrumentation*
  • Radiotherapy, Conformal / instrumentation*