Removal of a ruptured, detached, and entrapped angioplasty balloon after coronary stenting

J Invasive Cardiol. 2000 Feb;12(2):102-4.

Abstract

Entrapment of balloon catheters after coronary angioplasty or stent implantation is a complication during these procedures. We describe the removal of a ruptured angioplasty balloon which had become disconnected from the proximal shaft of the balloon catheter and entrapped after coronary stenting by using a 4 mm Amplatz "Goose Neck" Microsnare retrieval system (Microvena, White Bear Lake, Minnesota). The retrieval was accomplished without any complications. Retrieval systems should be on hand in all interventional catheterization laboratories.

Publication types

  • Case Reports

MeSH terms

  • Angioplasty, Balloon, Coronary / instrumentation*
  • Coronary Vessels*
  • Device Removal*
  • Equipment Failure
  • Humans
  • Male
  • Middle Aged
  • Stents*