Determination of finasteride in human plasma by liquid-liquid extraction and high-performance liquid chromatography

J Chromatogr B Biomed Sci Appl. 2000 Feb 11;738(2):305-10. doi: 10.1016/s0378-4347(99)00543-5.

Abstract

A high-performance liquid chromatographic method for the quantitation of finasteride in human plasma is presented. The method is based on liquid-liquid extraction with hexane-isoamylalcohol (98:2, v/v) and reversed-phase chromatography with spectrophotometric detection at 210 nm. The mobile phase consists of acetonitrile-15 mM potassium dihydrogenphosphate (40:60, v/v). Clobazam is used as the internal standard. The limit of quantitation is 4 ng/ml and the calibration curve is linear up to 300 ng/ml. Within-day and between-day precision expressed by relative standard deviation is less than 5% and inaccuracy does not exceed 8%. The assay was used for pharmacokinetic studies.

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Enzyme Inhibitors / blood*
  • Enzyme Inhibitors / pharmacokinetics
  • Finasteride / blood*
  • Finasteride / pharmacokinetics
  • Humans
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrophotometry, Ultraviolet