[Delivery of desflurane by a Tec 6 vaporizer at a concentration superior to that posted on the regulating valve]

Ann Fr Anesth Reanim. 1999 Jun;18(6):691-3. doi: 10.1016/s0750-7658(99)80159-2.
[Article in French]

Abstract

We report the case of a defective Tec 6 vaporizer (Datex-Ohmeda) which delivered higher concentrations of desflurane (end-tidal concentration: 3.5 vol%) than indicated by the vaporizer dial setting (2 vol%), at a fresh gas flow of 1 L.min-1. This condition was caused by a defective rotary control valve. Therefore it is essential to administer desflurane only in presence of a vapour analyser.