Clinical results of the CLARION magnetless cochlear implant

Ann Otol Rhinol Laryngol Suppl. 1999 Apr:177:22-6. doi: 10.1177/00034894991080s405.

Abstract

This paper reports initial results for the CLARION Multi-Strategy Cochlear Implant, presently under investigational study in Europe. A magnetless implantable cochlear stimulator (ICS) with an ear-mold-supported headpiece was designed in response to an increasing demand for a magnetic resonance imaging (MRI)-compatible cochlear implant. Surgical technique, accompanying magnetless headpiece, and MRI compatibility were evaluated in 11 deaf patients (ages 6 to 62 years) who were implanted with a magnetless Clarion implant. Because of the headset mechanics, the ICS was implanted closer behind the ear than a magnet-containing ICS. The ICS-MRI compatibility was investigated with 1.5- and 0.3-T MRI. Results showed that the surgery was relatively safe and easy to learn. The headset was stable and reliable. The MRI compatibility tests indicate that the ICS poses no contraindication for patients needing MRI. Overall, the results suggest that the Clarion magnetless cochlear implant is relatively safe and easy to implant, is MRI-compatible, and functions well with the ear-mold-supported headpiece.

Publication types

  • Clinical Trial

MeSH terms

  • Adult
  • Child
  • Cochlear Implantation / methods
  • Cochlear Implants*
  • Deafness / surgery
  • Equipment Design
  • Female
  • Humans
  • Magnetic Resonance Imaging*
  • Male
  • Middle Aged